8.3 Compensating Values to Match Reference Values (Load Compensation)
233
8
Chapter 8 Error Compens
ation
• Be sure to set the cable length before performing load compensation.
"8.4 Compensating Measurement Cable Errors (Cable Length Compensation)" (p. 245)
• Load compensation is enabled when the current measurement conditions match all com-
pensation conditions.
• If the current measurement frequency and compensation frequency do not match, an
error such as the following will be displayed on the Measurement screen, and compensa-
tion will be canceled.
• If a condition other than the compensation frequency does not match, compensation is
performed but an error like the following is displayed in the measurement screen.
• When the OPEN or SHORT compensation is valid, the load compensation is performed
for Z and
processed by the OPEN or SHORT compensation.
• In acquiring load compensation data (i.e., reference sample measurement), the OPEN/
SHORT compensation settings, that were defined before entry into the Load Compensa-
tion Screen, are valid.
• If the setting of the low Z high accuracy mode is changed, the compensation value
becomes invalid.
Summary of Contents for IM3533
Page 2: ......
Page 8: ...Contents vi...
Page 52: ...3 3 When TRANSFORMER Mode 44...
Page 156: ...4 5 Setting Application Settings 148...
Page 200: ...5 3 Application Settings 192...
Page 216: ...6 4 Setting Application Settings 208...
Page 270: ...9 4 Deleting a Panel 262...
Page 316: ...11 7 File and Folder Operations 308...
Page 338: ...12 6 Measurement Using a Computer 330...
Page 396: ...Appendix 13 Device Compliance Statement A26...
Page 399: ......
Page 400: ......