8.4 Compensating Measurement Cable Errors (Cable Length Compensation)
245
8
Chapter 8 Error Compens
ation
With high frequency measurement, the influence of the cable results in large measurement errors. Setting the
cable length enables you to reduce the measurement errors. Use a coaxial cable with 50
impedance.
8.4 Compensating Measurement Cable
Errors (Cable Length Compensation)
• If the cable length changes, repeat open, short, and load compensation.
• The guaranteed accuracy range varies with the cable length.
See:
Cable length coefficient (p. 345)
• When manufacturing your own cables, make sure that the cable length matches the length
set with the instrument. (IM3533: 1 m, IM3533-01: 1 m/ 2 m/ 4 m)
See:
"Points to pay attention to when making your own probe" (p. 32)
• When using the L2000, set cable length compensation to 1 m.
Procedure
Compensation Screen
1
LCR measurement Screen
2
Select the cable length to be used, and press
to confirm the selection.
Cable Length Compensation Setting
Select this when using a direct-coupled fix-
ture or the like.
Select this when the cable length is 1 m, 2m,
or 4 m.
3
Press
to close the setting screen.
The cable length settings for the IM3533-01 are as follows:
Select this when using a direct-coupled
fixture or the like.
Select this when the cable length is 1 m.
Select this when the cable length is 2 m.
Select this when the cable length is 4 m.
Cable Length Compensation Setting (IM3533-01only)
Summary of Contents for IM3533
Page 2: ......
Page 8: ...Contents vi...
Page 52: ...3 3 When TRANSFORMER Mode 44...
Page 156: ...4 5 Setting Application Settings 148...
Page 200: ...5 3 Application Settings 192...
Page 216: ...6 4 Setting Application Settings 208...
Page 270: ...9 4 Deleting a Panel 262...
Page 316: ...11 7 File and Folder Operations 308...
Page 338: ...12 6 Measurement Using a Computer 330...
Page 396: ...Appendix 13 Device Compliance Statement A26...
Page 399: ......