• The guaranteed accuracy range varies depending on the measurement conditions.
See:
Check the guaranteed accuracy range in "14.2 Measurement Range and Accuracy" (p. 342).
• The measurement range is determined according to the test range setting.If the display
for the measured value shows "
OVER FLOW
" or "
UNDER FLOW
", that means that mea-
surement cannot be performed using the currently set test range. Either you should set
AUTO ranging so as to select the most suitable test range automatically, or you should set
a more suitable test range manually.
• If the measurement range setting exceeds the range in the table above when the sweep
frequency is set, it is automatically changed to the maximum setting.
• When setting the sweep frequency, some ranges cannot be used depending on the fre-
quency range.
• 10 M
range: Up to 100.00 kHz
• 100 M
range: Up to 10.000 kHz
• With a test sample whose impedance varies with frequency, if the frequency is changed
during measurement using HOLD, measurement within the same range may not be possi-
ble. When this happens, change the measurement range.
• The test range setting is made according to the combination of the impedances of the
sample being tested and the test cables. Therefore it can happen that testing is not possi-
ble, if the test range is held with HOLD only upon the basis of the impedance of the sam-
ple under test. If this happens, you should change the test range, making reference to "8.1
Setting Open Circuit Compensation" (p. 215) and "8.2 Short Circuit Compensation" (p.
224).
4
Measurement Range Selection
Set the test range according to the combined impedance
value of the sample to be tested and the test cables.
Test
range
Accuracy guar-
anteed range
Auto ranging
range
100 M
8 M
to 200 M
8 M
or more
10 M
800k
to 100M
800 k
to 10 M
1 M
80k
to 10M
80 k
to 1 M
100 k
8 k
to
8 k
to 100k
10 k
800
to 100 k
800
to 10 k
1 k
80
to 10 k
80
to 1 k
100
8
to 100
8
to 100
10
800 m
to 10
800 m
to 10
1
80 m
to 1
80 m
to 1
100 m
10 m
to 100 m
0
to 100 m
To select the measurement range.
5
Press
to close the setting screen.
Summary of Contents for IM3533
Page 2: ......
Page 8: ...Contents vi...
Page 52: ...3 3 When TRANSFORMER Mode 44...
Page 156: ...4 5 Setting Application Settings 148...
Page 200: ...5 3 Application Settings 192...
Page 216: ...6 4 Setting Application Settings 208...
Page 270: ...9 4 Deleting a Panel 262...
Page 316: ...11 7 File and Folder Operations 308...
Page 338: ...12 6 Measurement Using a Computer 330...
Page 396: ...Appendix 13 Device Compliance Statement A26...
Page 399: ......
Page 400: ......