12.1 External Input/Output Connector and Signals
316
You can select rising or falling for the valid edge of a trigger.
See
"Enabling Trigger Input for during Measurement and Setting the Valid Edge of Trigger Input" (p. 136)
Signal function details
Input
TRIG
• When the trigger setting is the external trigger
, measurement is performed
once with the falling (ON) or rising (OFF) edge of the TRIG signal. The edge direction
can be set in the setting screen. (Initial value: Rising (ON))
See:
When the trigger source is set to the internal trigger
, trigger measurement is not
performed.
• You can set whether to enable or disable TRIG signal input during measurement (during
output of the EOM signal (HI)).
See:
LD0 to LD6
Selects the number of the panel to load.
If a trigger signal is input in external trigger mode, the selected panel is loaded and used for
measurement.
LD0, LD1
hen performing
measurement in TRANSFORMER mode, select LD0 and input
the trigger signal. When performing
measurement, select LD1 and input the trig-
ger signal.
LD-VALID
Inputs a negative logic signal from an external device so that the selected panel number is
recognized as valid.
After TRIG input, maintain a Low level until INDEX is outputted.
PIN No.
LD6
LD5
LD4
LD3
LD2
LD1
LD0
Panel 1
0
0
0
0
0
0
1
Panel 2
0
0
0
0
0
1
0
Panel 4
0
0
0
0
1
0
0
Panel 8
0
0
0
1
0
0
0
Panel 16
0
0
1
0
0
0
0
Panel 32
0
1
0
0
0
0
0
Panel 64
1
0
0
0
0
0
0
Panel 127
1
1
1
1
1
1
1
Panel128
0
0
0
0
0
0
0
0: (HIGH: 5 V to 24 V), 1: (LOW: 0 V to 0.9 V)
Summary of Contents for IM3533
Page 2: ......
Page 8: ...Contents vi...
Page 52: ...3 3 When TRANSFORMER Mode 44...
Page 156: ...4 5 Setting Application Settings 148...
Page 200: ...5 3 Application Settings 192...
Page 216: ...6 4 Setting Application Settings 208...
Page 270: ...9 4 Deleting a Panel 262...
Page 316: ...11 7 File and Folder Operations 308...
Page 338: ...12 6 Measurement Using a Computer 330...
Page 396: ...Appendix 13 Device Compliance Statement A26...
Page 399: ......
Page 400: ......