4.4 Judging Measurement Results
109
4
Chapte
r 4
LCR Function
Set the upper and lower limit values for two parameters and display up to 10 classifications of judgment
results.You can also output the judgment results to an external device.
BIN judgment starts with the first parameter for BIN1 and proceeds in order to BIN10, as described below.
The instrument will display the first BIN number for which the measurement value is judged to be within
the set judgment standard. If none of the BIN judgments is determined to apply, [
OUT OF BINS
] will be
displayed.
4.4.2
Classifying Measurement Results
(BIN Measurement)
About BIN function
Select the judgment mode for BIN measurement
before setting the judgment conditions (p. 100).
(When the measurement result is
within the BIN classification range)
(Front panel LEDs)
Lights red
(When the measure-
ment result is OUT-
OF-BINS)
Lights green
When BIN judgment
When BIN is not set
When does not match any BIN
By setting a series of increasingly lenient judgment standards as shown in the following dia-
gram, you can rank (sort) measurement elements.
Judgment reference
Summary of Contents for IM3533
Page 2: ......
Page 8: ...Contents vi...
Page 52: ...3 3 When TRANSFORMER Mode 44...
Page 156: ...4 5 Setting Application Settings 148...
Page 200: ...5 3 Application Settings 192...
Page 216: ...6 4 Setting Application Settings 208...
Page 270: ...9 4 Deleting a Panel 262...
Page 316: ...11 7 File and Folder Operations 308...
Page 338: ...12 6 Measurement Using a Computer 330...
Page 396: ...Appendix 13 Device Compliance Statement A26...
Page 399: ......
Page 400: ......