14.1 General Specifications
339
14
Chapter 14
S
pecifica
tions
Period of guaranteed
accuracy
1 year
Warm-up time
At least 60 minutes
Measurement time
Approx. 2.0 ms (1 kHz, FAST, no screen display)
Measurement speed
FAST, MED, SLOW, SLOW2
Terminal structure
4-terminal structure
Backup battery life
Approx. 10 years (25
C (77
F) reference value)
1.Basic Specifications
2.Function
Monitor functions
(1) Monitor voltage
• Monitor range
0.000 V to 5.000 V
• Monitor accuracy
±
10%rdg.
±
10 mV
(2) Monitor current
• Monitor range
0.000 mA to 100.0 mA
• Monitor accuracy
±
10%rdg.
±
10
A
Limit function
(1) Current limit (when V or CV)
• Limit range
0.01 mA to 100.0 mA
• Limit accuracy
±
10%rdg.
±
10
A
(2) Voltage limit (when CC set)
• Limit range
0.005 V to 5.000 V
• Limit accuracy
±
10%rdg.
±
10 mV
DC bias measurement
Superimposing a DC voltage and then performing measurement is possible.
DC voltage Normal mode: -5.00 V to 5.00 V (10 mV resolution)
Low Z high accuracy mode: -2.50 V to 2.50 V (10 mV resolution)
Generation accuracy:
±
10% of setting
±
(VAC 0.01 + 30 mV)
*VAC: AC signal voltage setting [V]
DC resistance
measurement
Settable when setting the Rdc measurement item.
The measurement condition when DC resistance measurement is settable to other than AC
measurement
• Measurement signal level: Normal mode: Fixed at 2 V
Low Z high accuracy mode: Fixed at 2 V
Occurrence accuracy:
±
10% of setting
±
20mV
• Measurement range, measurement speed, average, Adjustment delay, line frequency
• Temperature correction function: Converts data to referen
Reference temperature settings range ........-10
C to 99.9
C
Temperature coefficient setting range .........-99,999 ppm to 99,999 ppm
Temperature measurement
function
Temperature can be measured when the temperature T measurement parameter is set.
• Dedicated temperature probe: 9478 (option)
• Measurement range: -10.0
C to 99.9
C
• Guaranteed accuracy range: -10.0
C to 99.9
C
• Measurement accuracy: ±0.5% rdg ±1
C
• Ambient temperature: From 0
C to 18
C and 28
C to 40
C, add 0.02
C/
C to the mea-
surement accuracy.
• Sampling time: Approx. 640 ms
Average
1 to 256 (1 step)
Trigger function
An internal trigger or external trigger can be set.
Trigger delay
0 to 9.999 s (0.001s resolution)
BIN measurement
10 classifications for 2 items, OUT OF BINS
Absolute value setting,
% setting, % setting
Summary of Contents for IM3533
Page 2: ......
Page 8: ...Contents vi...
Page 52: ...3 3 When TRANSFORMER Mode 44...
Page 156: ...4 5 Setting Application Settings 148...
Page 200: ...5 3 Application Settings 192...
Page 216: ...6 4 Setting Application Settings 208...
Page 270: ...9 4 Deleting a Panel 262...
Page 316: ...11 7 File and Folder Operations 308...
Page 338: ...12 6 Measurement Using a Computer 330...
Page 396: ...Appendix 13 Device Compliance Statement A26...
Page 399: ......
Page 400: ......