5.2 Setting Basic Settings of Measurement
170
• The DC bias function is specifically for capacitor measurement. If it is used for resistor,
inductor, and other elements with low DC resistance, the following are likely.
• Normal measurement is not possible
• AUTO ranging is unable to determine a range.
• The DC bias function cannot be set during DC resistance measurement.
• The DC bias function cannot be set when
Rdc
measurement has been selected with the
:MEASure:ITEM
setting.
• When superimposing a DC voltage of
±
5 V (during low Z high accuracy mode operation,
±
2.5 V) or above, refer to "Appendix5.1 How to Supply a DC Bias Voltage"(p. A7).
• When superimposing a DC current on a coil or the like, refer to "Appendix5.2 How to Sup-
ply a DC Bias Current"(p. A9).
• If the total value for the measurement signal level (AC level setting value
×
+ DC bias
setting value) will become >
[V], the measurement signal cannot be raised any
higher. Reduce the AC level or DC bias value, and then configure the setting.
During low Z high accuracy mode operation, the AC level and DC bias value can be set
such that the total value is less than or equal to
[V].
Use
or
to set the DC bias.
5
4
Press
to close the setting screen.
• Settable range:
-5.00 V to 5.00 V (Normal mode)
-2.50 V to 2.50 V (Low Z high accuracy mode)
• If you make a mistake during input:
press
to cancel the input and start again.
DC Bias Setting
2
5 2
2.5 2
Summary of Contents for IM3533
Page 2: ......
Page 8: ...Contents vi...
Page 52: ...3 3 When TRANSFORMER Mode 44...
Page 156: ...4 5 Setting Application Settings 148...
Page 200: ...5 3 Application Settings 192...
Page 216: ...6 4 Setting Application Settings 208...
Page 270: ...9 4 Deleting a Panel 262...
Page 316: ...11 7 File and Folder Operations 308...
Page 338: ...12 6 Measurement Using a Computer 330...
Page 396: ...Appendix 13 Device Compliance Statement A26...
Page 399: ......
Page 400: ......