12.2 Timing Chart
320
Setting the INDEX and EOM output method
See
"4.5.7 Setting the EOM Output Method" (p. 137)
LCR Application Disk - communication commands (
:IO:EOM:MODE
)
Setting the pulse width for which low (on) EOM is held
See
"4.5.7 Setting the EOM Output Method" (p. 137)
LCR Application Disk - communication commands (
:IO:EOM:PULSe
)
• The shorter the measurement time, the shorter the time that INDEX and EOM are high
(off).
When the high (off) time is too short due to characteristics of the input circuit, the instru-
ment can be configured to maintain the low (on) state for a preset time once EOM
changes to low (on) before reverting the signal to high (off) after the completion of mea-
surement.
When trigger input is received at EOM:LOW and INDEX:LOW, the signal transitions to
high (off) when measurement starts.
(Start-of-measurement signal)
(Analog end of measurement signal)
(End of measurement signal)
(Analog end of measurement signal)
(End of measurement signal)
When HOLD
When PULSE
When HOLD
When PULSE
T: Pulse width