8.2 Short Circuit Compensation
227
8
Chapter 8 Error Compens
ation
The compensation values from last time are dis-
played in a confirmation screen.
(If compensation has never been performed, the
compensation values become 0.)
4
Short Compensation (All Compensation)
Check that the measurement cable is in a short-circuit
state.
Press .
When you want to limit the compensation
range:
Press
.
See
"Compensation range limitation function" (p. 218)
When you do not want to acquire the compen-
sation values:
Press .
The setting screen is redisplayed, and the compensation
values from last time become valid.
5
Performing Short Compensation (All Compensation)
Compensation starts.
Compensation value acquisition time:
Approximately 45 seconds
When you want to cancel compensation:
Press .
Compensation is cancelled and the compensation
screen is redisplayed.
(The short circuit compensation values from last time
remain.)
When unable to acquire normal compensa-
tion values:
When compensation failed:
When you want to disable open compensa-
tion data:
6
• If compensation ends normally, the effective resis-
tance and reactance are displayed.
• The possible compensation range is 1 k
or less
for impedance.
Use
or
to check the effective resistance
and reactance of each compensation point.
Short Compensation
(All Compensation)
Indicates the compensation results.
(Effective resistance, reactance)
Indicates the com-
pensation numbers.
Indicates the measure-
ment frequencies.
7
Press
to close the setting screen.
Summary of Contents for IM3533
Page 2: ......
Page 8: ...Contents vi...
Page 52: ...3 3 When TRANSFORMER Mode 44...
Page 156: ...4 5 Setting Application Settings 148...
Page 200: ...5 3 Application Settings 192...
Page 216: ...6 4 Setting Application Settings 208...
Page 270: ...9 4 Deleting a Panel 262...
Page 316: ...11 7 File and Folder Operations 308...
Page 338: ...12 6 Measurement Using a Computer 330...
Page 396: ...Appendix 13 Device Compliance Statement A26...
Page 399: ......
Page 400: ......