14.2 Measurement Range and Accuracy
345
14
Chapter 14
S
pecifica
tions
E
Cable length coefficient
The coefficient corresponding to the setting for measurement cable length is obtained from the measurement
cable length table and then multiplied by the basic accuracy.
*
IM3533 guaranteed accuracy range (frequency)
*
IM3533-01 guaranteed accuracy range (frequency)
0 m / 1 m / 2 m / 4 m: Up to 200 kHz (no limit)
F
DC bias coefficient
The coefficient corresponding to the setting for ON/OFF of DC bias is obtained from the DC bias coefficient
table and then multiplied by the basic accuracy.
G
Temperature coefficient
The coefficient corresponding to the operating temperature is obtained from the operating temperature coeffi-
cient table and then added to the basic accuracy.
0 m
1 m
2 m
4 m
Cable length
coefficient
IM3533
10 k
range or less
1
1.2
100 k
range or greater
1
1.2
IM3533-01
1
1.2
1.5
2
fm: Measurement frequency [kHz]
1.5
fm
100
----------
+
2
fm
50
-------
+
1.5
fm
20
-------
+
2
fm
10
-------
+
Cable length
10 k
range or less
100 k
range or greater
Guaranteed accuracy range
(frequency)
0 m
Up to 200 kHz
1 m
Up to 200 kHz
2 m
Up to 200 kHz
Up to 100 kHz
4 m
Up to 200 kHz
Up to 10 kHz
DC Bias Setting OFF
DC Bias Setting ON
DC bias
coefficient
1
2
0°C
t < 18°C, 28°C < t
40°C
18
°
C
t
28
°
C
Temperature
coefficient
1+0.1
×
|t - 23|
1
When the operating temperature (t) is 23
C
±
5
C, the coefficient is 1.
Summary of Contents for IM3533
Page 2: ......
Page 8: ...Contents vi...
Page 52: ...3 3 When TRANSFORMER Mode 44...
Page 156: ...4 5 Setting Application Settings 148...
Page 200: ...5 3 Application Settings 192...
Page 216: ...6 4 Setting Application Settings 208...
Page 270: ...9 4 Deleting a Panel 262...
Page 316: ...11 7 File and Folder Operations 308...
Page 338: ...12 6 Measurement Using a Computer 330...
Page 396: ...Appendix 13 Device Compliance Statement A26...
Page 399: ......
Page 400: ......