4.4 Judging Measurement Results
110
The comparator decision mode can be set as one of the following:
Deviation Percent (
%)
*2
Setting
HI
IN
IN
LO
upper limit value [
%]
reference value
lower limit value [
%]
Enter reference values and then set percentages corre-
sponding to the reference values as the upper limit and
lower limit
*1
values.
The measurement values are displayed in deviations (
%) from
the reference value.
Percent (%) Setting (p. 105)
HI
IN
IN
LO
upper limit value [%]
reference value
lower limit value [%]
Enter reference values and then set percentages corre-
sponding to the reference values as the upper limit and
lower limit
*1
values.
The measurement values displayed are the same as those of the
measurement parameters.
Absolute value (ABS) setting (p. 104)
HI
IN
LO
upper limit value
lower limit value
Set absolute values for the upper limit and lower limit values of
the measurement parameters.
The measurement values displayed are the same as those of the
measurement parameters.
• For more information about HI/IN/LO judgment procedures, see Page 103.
• If the power is turned off in BIN measurement mode, the mode will be BIN measurement
mode when the instrument starts the next time you turn the power on.
• For a BIN number that does not require a BIN judgment, set the upper and lower limit val-
ues to OFF.
• The measurement conditions that are used when normal measurement is performed are
inherited as is for the measurement conditions when BIN is performed.
• BIN measurement can be used even if only the upper or lower limit value has been set.
*
2: The following equation is used to calculate the
% value.
% =
|reference value|
measurement value-reference value
×100
*
1: The following equation is used to calculate the comparison upper limit value and comparison lower limit value.
(In the case of the comparison lower limit value, if a value that is lower than the reference value is set, the minus
(-) sign is required for the percentage setting value.)
Upper limit comparison value (Lower limit comparison value) = reference value + |reference value| ×
100
percentage set value
When only an upper limit value
has been set
HI
IN
Upper limit
value
When only a lower limit value
has been set
Lower limit
value
IN
LO
Summary of Contents for IM3533
Page 2: ......
Page 8: ...Contents vi...
Page 52: ...3 3 When TRANSFORMER Mode 44...
Page 156: ...4 5 Setting Application Settings 148...
Page 200: ...5 3 Application Settings 192...
Page 216: ...6 4 Setting Application Settings 208...
Page 270: ...9 4 Deleting a Panel 262...
Page 316: ...11 7 File and Folder Operations 308...
Page 338: ...12 6 Measurement Using a Computer 330...
Page 396: ...Appendix 13 Device Compliance Statement A26...
Page 399: ......
Page 400: ......