8.5 Calculating Values (Scaling)
248
•
If the same parameter is selected multiple times and a different compensation coefficient
is set for each of them, the compensation coefficient of the parameter with the smallest
number is used to perform scaling for all of the parameters of the parameter numbers.
(The compensation coefficients of the other parameter numbers become invalid.)
•
In the case of the following settings, scaling is performed using the compensation coeffi-
cient of parameter 1 for all Z of parameters 1, 2, and 4. (The compensation coefficients of
parameters 2 and 4 are invalid.)
6
Touch
and use the numeric keypad to
set compensation coefficient
B
in the same
way as for
A
. Then touch
to accept com-
pensation coefficient B.
Settable range: -9.99999G to 9.99999G
To return to the previous screen without making any
change to the set value, press the
key when the
screen is in the state with nothing being displayed (the
state after pressing the
key).
Changing the unit (
a/ f/ p/ n/
/ m/ None/ k/ M/ G
)
Step the units up.
Step the units down.
Compensation Coefficient Setting
7
Press
to return to the [Scaling Compensation Setting].
Reference value 1
Display Parameter Setting
Compensation Coefficient Setting
Parameter 1: Z
a = 1.500, b = 1.50000
Parameter 2: Z
a = 1.700, b = 2.50000
Parameter 3:
a = 0.700, b = 1.00000
Parameter 4: Z
a = 1.900, b = 3.50000
Summary of Contents for IM3533
Page 2: ......
Page 8: ...Contents vi...
Page 52: ...3 3 When TRANSFORMER Mode 44...
Page 156: ...4 5 Setting Application Settings 148...
Page 200: ...5 3 Application Settings 192...
Page 216: ...6 4 Setting Application Settings 208...
Page 270: ...9 4 Deleting a Panel 262...
Page 316: ...11 7 File and Folder Operations 308...
Page 338: ...12 6 Measurement Using a Computer 330...
Page 396: ...Appendix 13 Device Compliance Statement A26...
Page 399: ......
Page 400: ......