8.3 Compensating Values to Match Reference Values (Load Compensation)
243
8
Chapter 8 Error Compens
ation
You can clear all settings and repeat the configuration process from the compensation frequency setting
by touching
.
You can load the current measurement conditions (frequency, range, measurement signal mode and
value for the measurement signal level, and DC bias setting) as the load compensation conditions by
touching .
When You Want to Reset All Settings
When You Want to Use the Current Measurement Conditions as the
Load Compensation Conditions
Load Compensation Setting
Load Compensation Setting
When the measurement conditions are acquired with
,
Parameter to Use for Reference Value" (p. 241)) is initialized to Z-
.