5.2 Setting Basic Settings of Measurement
160
When measuring a sample whose impedance varies greatly with frequency or a sample whose characteristics
are unknown, you can have the instrument automatically select the optimal measurement range. Alternately,
you can perform high-speed measurement by fixing the range with the HOLD function.
There are the following two methods for setting the measurement range.
5.2.7 Setting the Measurement Range
1
Setting the method for determining the measurement range
(AUTO, HOLD)
The ranges are all defined in terms of impedance. Therefore, for a parameter other than
impedance, the value is obtained by calculating from the measured values of |Z| and
.
See
"Appendix1 Measurement Parameters and Calculation Formula"(p. A1)
Setting AUTO Ranging
AUTO
The most suitable test range is set automatically.
HOLD
The measurement range is fixed. The range is set manually.
Procedure
ANALYZER Measurement Screen
ANALYZER Basic Settings
1
Press
.
ANALYZER Basic Settings
2