10.4-34
Section 10: Calibrating the Model 9100: AC Current Function
Final Width = 215mm
Amplitude
Frequency
Table 10.4.6.3: Auxiliary AC Current Calibration Points
Output span
Calibration Targets
covered by
hardware
configuration
Default
Minimum
Maximum
Default
Minimum
Maximum
0.000mA to 320.0mA
300.000mA
180.000mA
320.000A
40Hz
35Hz
45Hz
(one range)
300.000mA
180.000mA
320.000A
150Hz
140Hz
160Hz
300.000mA
180.000mA
320.000A
1000Hz
900Hz
1100Hz
300.000mA
180.000mA
320.000A
3000Hz
2700Hz
3000Hz
0.3201A to 3.200A
3.00000A
1.80000A
3.20000A
40Hz
35Hz
45Hz
3.00000A
1.80000A
3.20000A
150Hz
140Hz
160Hz
3.00000A
1.80000A
3.20000A
1000Hz
900Hz
1100Hz
3.00000A
1.80000A
3.20000A
3000Hz
2700Hz
3000Hz
3.21 to 20.00A
10.0000A
7.00000A
10.5000A
40Hz
35Hz
45Hz
10.0000A
7.00000A
10.5000A
150Hz
140Hz
160Hz
10.0000A
7.00000A
10.5000A
1000Hz
900Hz
1100Hz
10.0000A
7.00000A
10.5000A
3000Hz
2700Hz
3000Hz