6-8
Section 6: 9100 System Operation — SCPI Language
Final Width = 215mm
6.4
Using the 9100 in a System
6.4.1
Addressing the 9100
6.4.1.1
Accessing the Bus Address
The instrument address can only be set manually; using the
Bus Address
menu, which
is accessed via the
Configuration
menus.
N.B.
A password is required for access to change the bus address.
6.4.1.2
Select 'Configuration' Mode
1. Press the Mode key on the right of the
front panel to obtain the 'Mode
Selection' menu screen:
2. Press the CONFIG screen key at the
center of the bottom row to progress
into 'Configuration' mode. The 9100
will transfer to the open 'Configuration'
menu screen:
6.4.1.3
Select 'MORE' Parameters
The bus address is one of the 'MORE' parameters. By trying to select 'MORE', the
'Configuration' password will be required.
Press the MORE screen key on the right of the bottom row. The 9100 will transfer
to the 'Password Entry' screen.
TODAYS DATE
TIME
PROC
MANUAL
CONFIG
CALIB
TEST
Mode Selection
Select required mode using softkeys
TODAYS DATE
TIME
M O R E
V I E W
Configuration
Present Settings:
Language
English
Power-up mode
Manual
Bus Address
1
Printer
NONE
Results Card
Disabled
Safety Voltage
. V
Border Line
7 . %
Scope option
Option 600
Crystal option
High acc
Ser. No. XXXXXX
Rev. XXX