Appendix D to Section 6: 9100 System Operation — Device Settings after
∗
RST
6-D3
Final Width = 215mm
6.D.3
∗
RST Settings Related to SCPI Commands
Keyword
Condition
CALibration
......................
Disabled
:SECure
:PASSword required to enter Calibration mode
OUTPut
[:STATe]
...................... OFF
:COMPensation(?) ............... OFF
:ISELection(?) ................. HIGHi
[SOURce]
:FUNCtion
[:SHAPe](?) ................. DC
:VOLTage
...................... Active
[:LEVEl]
[:IMMediate]
[:AMPLitude](?) ....... 1V
:HIGH(?) .............. Not applicable
:LOW(?) ............... Not applicable
:CURRent
...................... Inactive
:RESistance .................... Inactive
:UUT_I(?) ................... LOW
:CONDuctance ................... Inactive
:UUT_I(?) ................... LOW
:CAPacitance ................... Inactive
:UUT_I(?) ................... LOW
:FREQuency ..................... Inactive
:PHASe
...................... Inactive
[:ADJust](?) ................ 0
°
:INPut
[:STATe](?) .............. OFF
:OUTPut
[:STATe](?) .............. OFF
:PULSe
...................... Inactive
:TEMPerature ................... Inactive
:UNITs(?) ................... C
:SCALe(?) ................... TS68
:THERmocouple ............... Inactive
:TYPE(?) ................. K
:PRT
...................... Inactive
:TYPE(?) ................. PT385
:NRESistance(?) .......... 100
Ω
:UUT_I(?) ................ LOW
Keyword
Condition
STATus
:OPERation
[:EVENt]? .... No change
:ENABle(?) ... No change
:CONDition? .. Not applicable
:QUEStionable
[:EVENt]? .... No change
:ENABle(?) ... No change
:CONDition? .. Not applicable
:PRESet ......... Not applicable
SYSTem
:ERRor? ......... Not applicable
:DATE(?) ........ As previously set, updated
:TIME(?) ........ As previously set, updated
:SVOLtage(?) .... As previously set
:VERSion? ....... 1993