Section 10: Calibrating the Model 9100: Summary of Calibration Process
10.4-5
Final Width = 215mm
10.4.2.2
Sequencing Calibrations
The table below indicates the order in which the various Model 9100 functions should be
calibrated. Although it is not essential to calibrate all the functions indicated below at any
one time, functions higher in the list should be calibrated before those lower in the list.
If Resistance and Capacitance are being calibrated separately, the Resistance function
must be calibrated before the Capacitance function. Similarly if the Resistance
function's 'super' UUTi, 'high' UUTi and 'low' UUTi modes are being calibrated, some of
the 'super' UUTi hardware, then 'high' UUTi hardware configurations must be calibrated
before associated 'low' UUTi hardware configurations.
Sequence
Function
Reference
1
DC Voltage (Zero to 320V)
Section 10.4.3
2
AC Voltage (Zero to 320V)
Section 10.4.4
3
Resistance (Super UUTi)
Section 10.4.7
4
Resistance (High UUTi)
Section 10.4.7
5
Resistance (Low UUTi)
Section 10.4.7
6
DC Voltage (320V to 1050V)
Section 10.4.3
7
AC Voltage (320V to 1050V)
Section 10.4.4
8
Capacitance
Section 10.4.8
9
DC Current (Zero to 320mA)
Section 10.4.5
10
AC Current (Zero to 320mA)
Section 10.4.6
11
DC Current (0.32A to 20A)
Section 10.4.5
12
AC Current (0.32A to 20A)
Section 10.4.6
Table 10.4.2.1: Recommended Sequence of Calibrations