Appendix A to Section 8: 9100 Maintenance — Error Reporting Subsystem
8-A5
Final Width = 215mm
8.A.2.2.8
Device-Dependent Errors Reported only Remotely via the IEEE-488 Interface
The error list for remote operations, which are not reported on the front panel screen, is given below:
8.A.2.2.9
Device-Dependent Errors Reported both Locally and Remotely
Errors are reported both on the front panel screen and via the IEEE-488 interface. Note that the locally-presented error message
will not include the error number.
General
1001,"Overload of power amp"
1002,"Over temperature"
1003,"Over compliance"
1006,"Overload of HV power supply"
1008,"Overload of output clamp"
1009,"MULTIPLE analogue fail flags"
1010,"Softkey label too long"
1011,"Confirm with ON"
1014,"Power output not allowed with 9105
work-mat"
1020,"Unknown keycode"
5010,"Priority OFF received"
Calibration
4001,"Corrupt factors:- substitution"
4003,"Password incorrect"
4004,"Cal switch not enabled"
4005,"Factory password incorrect"
4006,"Correction block:- invalid"
4007,"Amplitude outside limits"
4008,"Cal is password protected"
4009,"Frequency outside limits"
4024,"Failed to save (act) factor"
4025,"Failed to save (tgt) factor"
4026,"Failed to save (frq) factor"
4027,"Failed to save (R-eqV) factor"
4028,"Limits: R-eqV"
4029,"Failed to save (act) R-dervd"
4030,"Failed to save (tgt) R-dervd"
4031,"Failed to save (act) C-ref"
4032,"Failed to save (tgt) C-ref"
4033,"Failed to save (frq) C-ref"
4034,"Failed to save (C-eqv) factor"
4035,"Failed to save (cjc) factor"
4051,"Cap meas no 1st reading"
4052,"Cap meas no 2nd reading"
4053,"Cap meas outside limits"
4055,"Corrupt cal factors"
4056,"Failed --------"
4057,"Corrupt selfcal factor"
4058,"Corrupt res. ref. factor"
4059,"Corrupt cold junc diff factor"
-300,"Device specific error"
-312,"PUD memory lost"
-315,"Selftest failed"
-9060,"Invalid V/div value"
-9061,"Illegal V/div, no. of div's
combination"
-9062,"-ve edge not available with 1M\352
load"
-9063,"Frequency too big for 1M\352 load"
-9064,"Amplitude too big for 50\352 load"
-9065,"Frequency too big for present
amplitude"
-9066,"Entered number exceeds limits"
-9067,"Search procedure - NO Test point"
-9068,"Search procedure - Function ID
Expected"
-9069,"Harmonic frequency cannot exceed
3000 Hz".
-9070,"Cannot change: phase too high".
-9070,"Cannot change: phase at
0\370"
-9071,"Cannot change: phase at
\36190\370"
-9072,"Cannot change: phase at
\361180\370"
-9073,"Cannot change: aux channel at
zero"
-330,"Configuration memory lost"
-350,"Queue overflow"
ALWAYS: record the total message content for possible use by the Service Center.
Continued Overleaf