Section 7: Model 9100: Specifications
7-1
Final Width = 215mm
SECTION 7
9100 SPECIFICATIONS
7.1.1
Line Power Supply
Voltage (single
φ
): 100V/120V/220V/240V
selectable from rear panel
Variation:
≤
±
10% Nominal Voltage
Line Frequency:
48Hz to 63Hz
Consumption:
450VA max
Power Fuses:
220/240V: T3.15A HBC, 250V, IEC127
100/120V: T5.0A HBC, 250V, IEC127
7.1.2
Mechanical
Dimensions:
Height:
3U
Width:
427mm (16.8 inches)
Depth:
460mm (18.1 inches)
Weight:
18.5kg (41lbs)
7.1.3
SAFETY:
Designed to UL1244, IEC348, IEC1010-1:
Pollution degree 2; installation category
II
;
Protection class
I
.
7.1
General
7.1.5
Environmental Conditions
Temperature:
Operating:
5
°
C to 40
°
C
Transit:
-20
°
C to 60
°
C <100hrs
Storage:
0
°
C to 50
°
C
Warm-up Time:
20 minutes
Max. Relative Humidity (non-condensing):
Operating:
+5
°
C to +30
°
C:
<90%,
+30
°
C to +40
°
C:
<75%,
Storage:
0
°
C to +50
°
C:
<95%,
Altitude
Operating:
0 to 2000m (6,562 ft)
Non-operating:
0 to 12,000m (40,000 ft)
Shock:
MIL-T-28800, type III, class 5, style E.
Vibration:
MIL-T-28800, type III, class 5, style E.
Enclosure:
MIL-T-28800, type III, class 5, style E.
EMC:
Designed to:
Generic Emissions:
EN50081.
Generic Immunity:
EN50082.
FCC Rules part 15 sub-part J class B.
7.1.6
The Meaning of 'Accuracy' when used in the
Function Accuracy Tables
Accuracy includes long-term stability, temperature coefficient,
linearity, load and line regulation and the traceability of factory
and National calibration standards. Nothing further needs to be
added to determine the Test Uncertainty Ratio over the instrument
under test.
N.B.
These specifications apply to both the Model 9100
output terminals and at the remote end of the Model 9105
lead kit unless otherwise stated.
7.1.4
Peak Terminal Voltages and Currents
Peak Volts
Peak
to Ground
Current
SHi
1500V
———
Hi
1500V
90mA
SLo
15V
———
Lo
15V
90mA
I+
10V
30A
I
-
———
30A
Aux Analog Output
15V
1.5A