10.4-62
Section 10: Calibrating the Model 9100: Continuity Function
Final Width = 215mm
10.4.10.4
Verification of Output Current Display
1. Connect a 1271/1281 or similar long-scale meter between the 9100 and the continuity
meter
2. Select the 1A range on the 1271
3. Set the 9100 output to zero ohms
4. Turn the 9100 output ON
5. Press the TEST button on the continuity meter and confirm that the current reading
displayed by the 9100 is within 1.5% of the current displayed on the reference meter.