Appendix C to Section 6: 9100 System Operation — IEEE 488.2 Common Commands and Queries
6-C11
Final Width = 215mm
6.C.12
Reset
∗
RST
will reset the instrument to a defined condition, stated for each
applicable command with the command's description, and listed
in Appendix D to Section 6.
The reset condition is not dependent on past-use history of the
instrument except as noted below:
∗
RST does not affect the following:
•
the selected address of the instrument;
•
calibration data that affect specifications;
•
SRQ mask conditions;
•
the state of the IEEE 488.1 interface.
Execution Errors:
None.
Power On and Reset Conditions
Not applicable.
6.C.13
Service Request Enable
This Status Byte data structure conforms to the IEEE 488.2 standard requirements for this structure.
∗
SRE enables the standard and user-defined summary bits in the service request byte, which will generate a service request. Refer
to Section 6, Subsection 6.5.
Execution Errors:
None.
Power On and Reset Conditions
Not applicable.
Nrf
is a Decimal Numeric Data Element representing an
integer decimal value equivalent to the Hex value required
to enable the appropriate bits in this 8-bit register. The
detail definition is contained in the IEEE 488.2 document.
Note that numbers will be rounded to an integer.
RST
*
Nrf
phs
SRE
*