10.4-54
Section 10: Calibrating the Model 9100: Insulation Resistance Function
Final Width = 215mm
10.4.9.4
Calibration Setup
1. Connections
Ensure that the 9100 is connected to the insulation tester as shown
in Fig. 10.4.9.1 and that both instruments are powered on and
warmed up.
2. 9100
Ensure that the 9100 is in STANDARD CAL mode and then select
the Insulation Resistance function by pressing the
AUX
key on the
right of the front panel followed by the softkey.
Ω
Output span covered by Suitable output Recommended Minimum Maximum Factor
hardware
to select h/w
Number
configuration
configuration
100.0K 299.9K
200.0K
110.0K
100.0K
120.0K
1
270.0K
240.0K
299.9K
2
300.0K
2.999M
2.000M
330.0K
300.0K
360.0K
1
1.000M
0.900M
1.100M
2
3.000M
29.99M
20.00M
3.300M
3.000M
3.600M
1
10.00M
9.000M
11.00M
2
30.00M
299.9M
200.0M
33.00M
30.00M
36.00M
1
100.0M
90.00M
110.0M
2
300.0M
2.000G
0.600G
660.0M
600.0M
720.0M
1
1.800G
1.600G
2.000G
2
Table 10.4.9.1: Insulation Resistance Hardware Configurations and Calibration Points