Section 8: Model 9100 — Routine Maintenance and Test
8-15
Final Width = 215mm
TODAYS DATE
TIME
8.3.3
Selftest at Power-On
N.B.
Certain catastrophic 'Fatal System Errors' may cause the display to flash on and
off, at the point of setting power on. In this case, immediately switch Power Off
and report the fault to your Service Center.
The first normal action at power on is to show the
Fluke
logo, and then the 9100 will
run a fast selftest. The user is requested to wait until the selftest is finished:
Please wait approx. 1 minute while
internal checks are performed.
If no failures are found, the 9100 will revert to the default power-up mode, which is
either Manual mode or Procedure mode.
If failures are encountered (including other 'Fatal System Errors' — refer to Appendix A
to this Section 8), the 9100 will lapse into Test mode at the following stage:
E X I T
Selftest.
Selected test has FAILED.
Number of failures: XX
Use the softkeys to view the results.
P R I N T
VIEW
FAILS
Subsequent action to view the failures and print the results follows as for a Fast selftest
which has run to completion (refer to paras 8.3.2.2/3).
EXIT
will return to the '
Select selftest
' menu screen, where, to return to the '
Mode
Selection
' menu screen, merely press the front panel
Mode
key.
can be used to print out the results of the power-on Fast test.
Refer also to Sub-section 8.4.