7-28
Section 7: Model 9100: Specifications
Final Width = 215mm
7.19.4 Auxiliary AC Voltage (Representing AC Current Output) Accuracy
NOTES: Improved specs are available at intermediate points depending on scaling factor.
Auxiliary AC voltage is available as a simulation of the output of a resistive shunt, and can therefore be expressed as being equivalent to Amps.
Also directly available as a voltage harmonic output. When used as a harmonic with n>1, add 0.04% to percentage of output specifications.
7.19.5 Other Voltage Waveshapes
All subject to 65Hz/1kHz maximum as defined in main Model 9100 specification tables.
Note that all waveshapes are specified to 4.525 pk only (4 ranges; equivalent to the 3.2V sinewave range peak).
Functionality above this limit is available but unspecified.
The 4.525 mV pk range is allowed to function down to zero.
Specifications as main 9100 ranges.
Output Current
Frequency
Accuracy
Compliance Current
Band
±
(% Floor)
mA (O/P Impedance)
0.32A to 3.20A
10 - 3K
0.12% + 0.01A
10
Ω
3.21A to 32.00A
10 - 3K
0.12% + 0.01A
10
Ω
32.1 to 320.0A @
≥
1mV/Amp
10 - 3K
0.04% + 0.02A
<20mA
321A to 3.2KA @
≥
1mV/Amp
10 - 3K
0.04% + 0.2A
<20mA
0 to 160.0A @ 0.2mV to 1mV/Amp
10 - 3K
0.04% + 0.1A
<20mA
160A to 3.2kA @ 0.2mV to 1mV/Amp
10 - 3K
0.04% + 1A
<20mA
>3.2kA
10 - 3K
0.06% + 2A
<7mA @ 7V
7.19.6 DC Current Accuracy
Output Current
Accuracy
±
(% Floor)
0 to 320.0mA
0.016 + 9.6
µ
A
0.320A to 3.200A
0.06 + 118
µ
A
3.20A to 10.50A
0.055 + 940
µ
A
10.50A to 20.00A
0.055 + 4.5mA
3.200 to 32.00A, 10t coil
0.26 + 1.18mA
32.00 to 200.0A, 10t coil
0.255 + 45mA
16.00 to 160.0A, 50t coil
0.26 + 5.9mA
160.0 to 1000A, 50t coil
0.255 + 225mA