10.4-52
Section 10: Calibrating the Model 9100: Insulation Resistance Function
Final Width = 215mm
10.4.9.2
Calibration Equipment Requirements
• A set of traceably calibrated standard resistors at values between the minimum and
maximum targets defined in the following table, capable of withstanding at least
1000V DC for a prolonged period.
• A 3
1
/
2
digit (or better) insulation test meter (for example, a Yokogawa 2426). This is
used to make transfer measurements between the standard resistors and the Model
9100.