Section 7: Model 9100: Specifications
7-9
Final Width = 215mm
7.6
AC Current Specifications
7.6.1
AC Current Accuracy (Sinusoidal Waveshape)
Current
Frequency
Accuracy †
Compliance
Compliance
Total
Compliance
Absolute
Output
Band
[2]
Voltage
Voltage
Harmonic
Error (A/V)
Resolution
(Hz)
±
(% Floor)
(VRMS at 9100
(VRMS at 9105
Distortion
for
1 Year - Tcal
[1]
±
5
°
C
terminals)
Lead End)
(% Output)
Vc>0.5VRMS
000.000
µ
A - 032.000
µ
A
10 - 3k
0.07 + 900nA
4V
4V
0.10
60nA/V
1nA
3k - 10k
0.10 + 1.8
µ
A
4V
4V
0.25
600nA/V
1nA
10k - 20k
0.20 + 6.0
µ
A
4V
4V
0.40
2.4
µ
A/V
1nA
20k - 30k
0.25 + 9.0
µ
A
4V
4V
0.60
5.4
µ
A/V
1nA
032.001
µ
A - 320.000
µ
A
10 - 3k
0.07 + 300nA
4V
4V
0.10
60nA/V
1nA
3k - 10k
0.10 + 600nA
4V
4V
0.25
600nA/V
1nA
10k - 20k
0.20 + 2.0
µ
A
4V
4V
0.40
2.4
µ
A/V
1nA
20k - 30k
0.25 + 3.0
µ
A
4V
4V
0.60
5.4
µ
A/V
1nA
0.32001mA - 3.20000mA
10 - 3k
0.07 + 300nA
4V
4V
0.10
60nA/V
10nA
3k - 10k
0.10 + 600nA
4V
4V
0.25
600nA/V
10nA
10k - 20k
0.20 + 2.0
µ
A
4V
4V
0.40
2.4
µ
A/V
10nA
20k - 30k
0.25 + 3.0
µ
A
4V
4V
0.60
5.4
µ
A/V
10nA
03.2001mA - 32.0000mA
10 - 3k
0.07 + 3.2
µ
A
4V
4V
0.10
0.5
µ
A/V
100nA
3k - 10k
0.10 + 6.4
µ
A
4V
4V
0.25
4
µ
A/V
100nA
10k - 20k
0.20 + 12.8
µ
A
4V
4V
0.40
15
µ
A/V
100nA
20k - 30k
0.25 + 22.4
µ
A
4V
4V
0.60
32
µ
A/V
100nA
032.001mA - 320.000mA
10 - 3k
0.08 + 32.0
µ
A
4V
4V
0.10
2
µ
A/V
1
µ
A
3k - 10k
0.10 + 48.0
µ
A
4V
4V
0.25
4
µ
A/V
1
µ
A
10k - 20k
0.20 + 64.0
µ
A
4V
4V
0.40
15
µ
A/V
1
µ
A
20k - 30k
0.25 + 96.0
µ
A
4V
4V
0.60
35
µ
A/V
1
µ
A
0.32001A - 3.20000A
10 - 3k
0.10 + 480
µ
A
2.5V
2.4V
0.20
90
µ
A/V
10
µ
A
3k - 10k
0.25 + 2.56mA
2.5V
2.4V
1.10
600
µ
A/V
10
µ
A
03.2001A - 10.5000A
10 - 3k
0.20 + 3.0mA
2.5V
2.3V
0.20
0.3mA/V
100
µ
A
3k - 10k
0.50 + 10.0mA
2.2V
2.0V
1.10
2.1mA/V
100
µ
A
10.5001A - 20.0000A
*
10 - 3k
0.20 + 6.9mA
2.5V
♦
2.2V
♦
0.30
0.3mA/V
100
µ
A
3k - 10k
0.50 + 23.0mA
2.1V
1.7V
1.50
2.1mA/V
100
µ
A
03.2001A - 32.0000A
♥
10 - 100
0.20 + 5.5mA
2.5V
2.5V
0.15
---
100
µ
A
100 - 440
0.78 + 27mA
2.5V
2.5V
0.50
---
100
µ
A
032.001A - 200.000A
*
♥
10 - 100
0.21 + 90mA
2.5V
♦
2.3V
♦
0.15
---
1mA
100 - 440
0.67 + 0.25A
2.5V
2.3V
0.50
---
1mA
016.001A - 160.000A
♣
10 - 100
0.20 + 28mA
2.5V
2.5V
0.15
---
1mA
0160.01A - 1000.00A
*
♣
10 - 100
♠
0.21 + 0.45A
2.5V
♦
2.3V
♦
0.15
---
10mA
(More ACI Specifications overleaf)
NOTES: [1]
Tcal = temperature at calibration. Factory calibration temperature = 23
°
C.
[2]
Frequency Accuracy: 25ppm of output frequency.
∗
= With output 'ON', maximum duty cycle of (>0.525FS :
≤
0.525FS) is (1 : 4).
Continuous output >0.525FS will automatically reduce to <0.525FS after 2 Minutes.
† = Total uncertainty includes compliance errors for Voltage
≤
0.5VRMS.
Above 0.5V, add appropriate compliance error, except for Outputs marked
♥
and
♣
.
♥
= Refers to accuracy at 9100 output terminals. With Option 200 10 turn coil connected, then at the output from the coil, add
±
0.2% of output from
coil for uncertainty of coil.
♣
= Refers to accuracy at 9100 output terminals. With Option 200 50 turn coil connected, then at the output from the coil, add
±
0.2% of output from
coil for uncertainty of coil.
♦
= For frequencies <40Hz, compliance voltage is reduced by 0.5V RMS.
♠
= These coils have been designed for optimum accuracy and inductance for use with the Model 9100. With some Hall effect clamp meters the
increase in inductance, due to the current clamp design, will limit the obtainable 9100 Current/Hertz profile. In some cases, 1000A cannot be
reached at higher frequency.