Section 10: Calibrating the Model 9100: Capacitance Function
10.4-49
Final Width = 215mm
10.4.8.7
Calibration of 'UUTi Super' Source Current Hardware Configurations
For calibration of UUTi Super source current hardware configurations use Table 10.4.8.2.
Output span covered
Suitable output to
Calibration Targets
Factor
by hardware configuration
select hardware
Number
configuration
Recommended
Minimum
Maximum
40.001
µ
F to 400.000
µ
F
100
µ
F
60.000
µ
F
41.000
µ
F
80.000
µ
F
1
190.000
µ
F
250.000
µ
F
390.000
µ
F
2
0.40001mF to 4.00000mF
1mF
0.60000mF
0.41000mF
0.80000mF
1
1.90000mF
2.50000mF
3.90000mF
2
4.0001mF to 40.0000mF
10mF
6.0000mF
4.1000mF
8.0000mF
1
19.0000mF
25.0000mF
39.0000mF
2
Table 10.4.8.2: Capacitance UUTi Super Hardware Configurations and Calibration Points