10.4-16
Section 10: Calibrating the Model 9100: AC Voltage Function
10.4.4.4
Calibration Setup
1. Connections
Ensure that the 9100 is connected to the Standards DMM as shown in Fig. 10.4.4.1, and that both instruments
are powered on and warmed up.
2. 9100
a. Ensure that the 9100 is in STANDARD CAL mode and then select the AC Voltage function by pressing
the 'V' key on the right of the front panel, followed by pressing the '
V' screen key adjacent to the display.
b. Ensure that Sinusoidal waveshape is selected, and that the Phase facility is deselected (
∆Φ
label dark
lettering on light background).
Output span
Output
Calibration Targets
Factor
covered by
to select
Number
hardware
hardware
configuration
config.
Default
Minimum
Maximum
Default
Minimum
Maximum
320.01V
500V
750.00V
700.00V
1050.00V
40.00Hz
40.00Hz
75.00Hz
1
to
750.00V
700.00V
1050.00V
200.00Hz
150.00Hz
250.00Hz
2
1050.00V
750.00V
700.00V
1050.00V
10.000kHz
3.0000kHz
20.000kHz
3
330.00V
320.10V
350.00V
30.000kHz
25.000kHz
30.000kHz
4
Amplitude
Frequency
Table 10.4.4.2: AC Voltage (320V to 1050V) Hardware Configurations and Calibration Points
Output span
Output
Calibration Targets
Factor
covered by
to select
Number
hardware
hardware
configuration
config.
Default
Minimum
Maximum
Default
Minimum
Maximum
000.000mV
100mV
300.000mV
180.000mV
320.000mV
1.0000kHz
150.00Hz
5.000kHz
1
to
300.000mV
180.000mV
320.000mV
30.000kHz
15.000kHz
45.00kHz
2
320.000mV
300.000mV
180.000mV
320.000mV
100.00kHz
65.00kHz
100.00kHz
3
0.32001V
1V
3.00000V
1.80000V
3.20000V
1.0000kHz
150.00Hz
5.000kHz
1
to
3.00000V
1.80000V
3.20000V
30.000kHz
15.000kHz
45.00kHz
2
3.20000V
3.00000V
1.80000V
3.20000V
100.00kHz
65.00kHz
100.00kHz
3
3.2001V
10V
30.0000V
18.0000V
32.0000V
1.0000kHz
150.00Hz
5.000kHz
1
to
30.0000V
18.0000V
32.0000V
30.000kHz
15.000kHz
45.00kHz
2
32.0000V
30.0000V
18.0000V
32.0000V
100.00kHz
65.00kHz
100.00kHz
3
32.001V
100V
100.000V
70.000V
105.000V
1.0000kHz
150.00Hz
5.000kHz
1
to
100.000V
70.000V
105.000V
30.000kHz
15.000kHz
45.00kHz
2
105.000V
100.000V
70.000V
105.000V
100.00kHz
65.00kHz
100.00kHz
3
105.001V
150V
300.000V
180.000V
320.000V
40.00Hz
40.00Hz
75.00Hz
1
to
300.000V
180.000V
320.000V
200.00Hz
150.00Hz
250.00Hz
2
320.000V
300.000V
180.000V
320.000V
10.000kHz
3.0000kHz
20.000kHz
3
300.000V
180.000V
320.000V
30.000kHz
25.000kHz
30.000kHz
4
Amplitude
Frequency
Table 10.4.4.1: AC Voltage (Zero to 320V) Hardware Configurations and Calibration Points