8.3.4.4
Display Checks
Display checks are initiated by pressing the DISPLAY key on the 'Select test' menu
screen:
D I S P L A Y
Selecting 'DISPLAY' transfers to the 'Display test.' screen. This invites a user to use the
front panel'tab' (
) key to move the dark band down the screen, up to the top, and down
again, thus testing all the display elements on the screen:
Display test.
Use the tab key to advance the band
and the enter key to exit the test.
The screen text will shift automatically so as not to obscure the band as it jumps back
to the top of the screen. If there are elements of the screen which do not show light in
the light background or dark in the dark band as it is advanced down the screen, then this
implies a failure. Rectification will require access to the internal circuitry, so no further
user action is recommended, except to report the result to your
Fluke
Service Center.
The '
Enter
' Key (
↵
)
returns to the Interface '
Select test
' menu screen.