Section 10: Calibrating the Model 9100: AC Voltage Function
10.4-15
sHi
sLo
Hi
Lo
I
+
I-
1500Vpk
max
15Vpk
max
15Vpk
max
HIGH VOLTAGE
DANGER
!
!
I+
I-
Hi
Lo
Reset
Guard
AAA
AAA
AAA
AAA
AAA
AAA
AAA
AAA
AAA
AAA
AAA
AAA
AAA
AAA
AAA
!
Coaxial Cable
e.g. 1513
Standards DMM
Ω
Guard
Fig 10.4.4.1 AC Voltage Calibration — Interconnections
10.4.4.3
Interconnections