10.4-8
Section 10: Calibrating the Model 9100: DC Voltage Function
Final Width = 215mm
10.4.3.4
Calibration Setup
1. Connections
Ensure that the 9100 is connected to the Standards DMM as shown in Fig. 10.4.3.1, and that both instruments
are powered on and warmed up.
2. 9100
Ensure that the 9100 is in STANDARD CAL mode and then select the DC Voltage function by pressing the
'
V
' key on the right of the front panel.
Output span covered
Suitable output to
Calibration Targets
Factor
by hardware configuration
select hardware
Number
configuration
Default
Minimum
Maximum
000.000mV to
±
320.000mV
100mV
000.000mV
-0.080mV
10.000mV
1
+300.000mV
180.000mV
320.000mV
2
±
0.32001V to
±
3.20000V
1V
-3.00000V
-3.20000V
-1.80000V
1
+3.00000V
+1.80000V
+3.20000V
2
±
3.2001V to
±
32.0000V
10V
-30.0000V
-32.0000V
-18.0000V
1
+30.0000V
+18.0000V
+32.0000V
2
±
32.001V to
±
320.000V
100V
-300.000V
-320.000V
-180.000V
1
+300.000V
+180.000V
+320.000V
2
Output span covered
Suitable output to
Calibration Targets
Factor
by hardware configuration
select hardware
Number
configuration
Default
Minimum
Maximum
±
320.01V to
±
1050.00V
1000V
-1000.00V
-1050.00V
-700.00V
1
+1000.00V
+700.00V
+1050.00V
2
Table 10.4.3.2: DC Voltage (320V to 1050V) Hardware Configurations and Calibration Points
Table 10.4.3.1: DC Voltage (Zero to 320V) Hardware Configurations and Calibration Points