10.3-10
Section 10: Calibrating the Model 9100: Standard Calibration — Basic Sequences
Final Width = 215mm
10.3.4.2
Changing the Target Value
As mentioned earlier, in addition to selecting either:
•
the previously-saved calibration target values or
•
default calibration targets,
a new target value can be defined. This is achieved on the Calibrate screen, for the
selected factor, by editing the target value.
To do this, first transfer cursor control to the target value displayed in the lower left of
the screen by pressing the Tab
key. (Note: the Tab
key moves the cursor between
alternate fields, so that re-pressing it will return cursor control to the output amplitude
control display.)
With the cursor on the displayed target value, you can now alter the target value using
digit or direct edit employed in normal operation — (Model 9100 Universal Calibration
System User's Handbook, Volume 1, Operation, Section 3, sub-section 3.4).
Note that although the x10,
÷
10 and
±
softkeys remain active, pressing them will always
result in the error message:
Target too big
or
Target too small
being displayed, as they would move the value out of the permissible target range.
Pressing the ZERO softkey will also produce this error message, unless the default target
value is already zero (as would be the case, for example, for Factor 1 of the DC Voltage,
0V to
±
320.000mV hardware configuration).