8-20
Section 8: Model 9100 — Routine Maintenance and Test
Final Width = 215mm
TODAYS DATE
TIME
TODAYS DATE
TIME
8.3.4.5
Memory Card Checks
Memory Card checks are initiated by pressing the MEMORY CARD key on the 'Select
test' menu screen:
E X I T
Selftest.
Select test using softkeys
D I S P L A Y
M E M O R Y
T R A C K E R
P R I N T E R
K E Y B R D
D I S P L A Y
M E M O R Y
C A R D
Pressing the MEMORY CARD key on the 'Select test' menu screen transfers to the 'Card
slot test.' screen. This invites a user to select the memory card slot to be tested, and
presents an 'Overwrite' warning:
Card slot test.
Select the card slot to be
tested using the soft keys.
WARNING!!
Card data will be overwritten.
E X I T
P R O C
R E S U L T S
PROC
selects the
PROCEDURE
slot for testing.
RESULTS
selects the
RESULTS
slot for testing.
EXIT
returns to the Interface '
Select test
' menu screen.
Either slot can be used for the test. If one particular slot is suspect then that, naturally,
will be the one to choose.
The warning gives notice that the inserted memory card will be overwritten by this test.
This is because reformatting is required to perform the checks and the card inserted into
the slot will be reformatted as a Results card.
It is not possible to format a Procedure card using this test, as procedures are written onto
cards by a different process (
Fluke
Model 9010).