8-14
Section 8: Model 9100 — Routine Maintenance and Test
Final Width = 215mm
TODAYS DATE
TIME
8.3.2.3
Viewing the Test Results
By pressing the
VIEW FAILS
screen key, each of the failed tests can be viewed in turn,
on a screen which shows the test number (pathway), averaged measured value, upper
and lower limits, and the values of the two measurements which were used to derive the
'Measured value'. A brief description of the test is also given.
The screen for viewing the test results is shown below. This can also be used when a
test has been aborted.
E X I T
Selftest.
P R I N T
NEXT
FAILURE
Test no. XXX.XXX FAIL.
Measured value: XXXXXXX
Upper Limit:
XXXXXXX
Lower Limit:
XXXXXXX
Measurement A:
XXXXXXX
Measurement B:
XXXXXXX
(Description
of test)
N.B.
If the cause of failure is not immediately obvious, and it is intended to consult
your
Fluke
Service Center, please ensure that you either: copy the details
from the screen for all the reported failures, or: print out the results. No second
viewing of the same failure is allowed, although all the test results remain
available for printing.
NEXT FAILURE
Once the details of the first failure have been noted, the next failure in the list can be
viewed by pressing the
NEXT FAILURE
screen key. Only one pass through the list of
failures is allowed. Once the last failure in the list is on the screen, and the
NEXT
FAILURE
screen key is pressed, the following error message will appear in the top right
of the screen:
No more failures
to view
Pressing the
NEXT FAILURE
screen key will have no further effect. The user must either
exit back to the 'Select selftest' screen, or print out the results.
8.3.2.4
Printing the Test Results
The
screen key is present on all screens after the test has run to completion or has
been aborted. Pressing the
key will print out all the available results.
Printing will only be possible if a suitable printer is set up, connected and on line. Refer
to Sub-section 8.4.