3.4.9
Rosemount 5900C in a F
OUNDATION
™
Fieldbus system
The Rosemount 5900C Radar Level Gauge supports the F
OUNDATION
Fieldbus (FF)
technology and lets you integrate it into an existing FF network. As long as the power
supply meets certain requirements (see
) the Rosemount
will be able to operate as any other FF device.
Figure 3-28: Example of an I.S. F
OUNDATION
Fieldbus System
$
%
&
'
(
)
A. I.S. Power Supply
B. Trunk
C. Rosemount 2230 Display
D. Segment Coupler
E. Rosemount 644 Temperature Transmitter
F. Rosemount 5900C Radar Level Gauge
Ensure that the power supply is able to provide the total current needed for all the
connected devices. See
and
for further information.
Ensure that:
•
the Rosemount 5900C and other devices connected to the F
OUNDATION
Fieldbus (FF)
system are compliant with the FISCO or Entity parameters of the power supply.
•
the short circuit protection of the Segment Coupler
consumption of the connected devices.
(7)
See
for Rosemount 5900C approval information
(8)
See the Rosemount 2410 Reference Manual (Document No. 300530EN) for more information on the Segment Coupler.
Reference Manual
Installation
00809-0100-5901
September 2019
Rosemount 5900C Radar Level Gauge
109
Summary of Contents for Rosemount 5900C
Page 1: ...Reference Manual 00809 0100 5901 Rev DA September 2019 Rosemount 5900C Radar Level Gauge ...
Page 6: ...Contents Reference Manual September 2019 00809 0100 5901 vi Reference Manual ...
Page 12: ...Introduction Reference Manual September 2019 00809 0100 5901 12 Reference Manual ...
Page 116: ...Installation Reference Manual September 2019 00809 0100 5901 116 Reference Manual ...
Page 204: ...Configuration Reference Manual September 2019 00809 0100 5901 204 Reference Manual ...
Page 208: ...Operation Reference Manual September 2019 00809 0100 5901 208 Reference Manual ...
Page 290: ...Product Certifications Reference Manual September 2019 00809 0100 5901 290 Reference Manual ...