8. Check that the line fits well to the measurement points. If a point deviates
significantly from the line, it can be excluded from the calculations. Open the
Calibration Data
window (click the
Calibration Data
button) and uncheck the
corresponding check box in the
Enable
column.
9. Click the
Write new calibration data to RTG
button to save the current calibration
data to the level gauge database registers.
Note
By clicking the
Write new calibration data to RTG
button, the Level values in the
Calibration Data
window are recalculated and the old Calibration Data is replaced.
Now you can check the calibration result in the
Calibrate
window again:
Note that all measured values are adjusted according to the calculated Calibration
Distance and Correction Factor. In the
Calibration Data
window you can also see that
the level values measured by the Rosemount 5900C gauge are adjusted. Of course,
the hand dipped levels are unaltered.
Note
Do not change the Calibration Distance in the
Properties/Tank Geometry
window
when calibration is finished.
Configuration
Reference Manual
September 2019
00809-0100-5901
154
Reference Manual
Summary of Contents for Rosemount 5900C
Page 1: ...Reference Manual 00809 0100 5901 Rev DA September 2019 Rosemount 5900C Radar Level Gauge ...
Page 6: ...Contents Reference Manual September 2019 00809 0100 5901 vi Reference Manual ...
Page 12: ...Introduction Reference Manual September 2019 00809 0100 5901 12 Reference Manual ...
Page 116: ...Installation Reference Manual September 2019 00809 0100 5901 116 Reference Manual ...
Page 204: ...Configuration Reference Manual September 2019 00809 0100 5901 204 Reference Manual ...
Page 208: ...Operation Reference Manual September 2019 00809 0100 5901 208 Reference Manual ...
Page 290: ...Product Certifications Reference Manual September 2019 00809 0100 5901 290 Reference Manual ...