17. Select the
Gas Correction
tab.
18. Choose the appropriate correction method for the product in the tank:
Option
Description
Air Correction
This method should only be used when there is no vapor in
the tank, i.e. when the tank is empty and contains air only. It
is used in the initial step when calibrating the Rosemount
5900C.
One known gas
This method may be used when there is only one gas type in
the tank. It provides the highest accuracy among the
different correction methods. Note that even small amounts
of another gas reduces the accuracy.
One or more
unknown gases
Use this method for hydrocarbons, for example Propane/
Buthane, when the exact mixture is not known.
Two gases with
unknown mixratio
This method is suitable for a mixture of two gases even if the
mixratio is not known.
One or more known
gases with known
mixratio
This method may be used when there is a well known
mixture of up to 4 products in the tank.
Now the Rosemount 5900C level gauge is ready to measure the product level when
the tank is put into operation.
F
OUNDATION
Fieldbus parameters:
TRANSDUCER 1500>LPG_CORRECTION_METHOD
TRANSDUCER 1500>LPG_NUMBER_OF_GASSES
TRANSDUCER 1500>LPG_GAS_TYPE1, TRANSDUCER 1500>LPG_GAS_PERC1
Configuration
Reference Manual
September 2019
00809-0100-5901
202
Reference Manual
Summary of Contents for Rosemount 5900C
Page 1: ...Reference Manual 00809 0100 5901 Rev DA September 2019 Rosemount 5900C Radar Level Gauge ...
Page 6: ...Contents Reference Manual September 2019 00809 0100 5901 vi Reference Manual ...
Page 12: ...Introduction Reference Manual September 2019 00809 0100 5901 12 Reference Manual ...
Page 116: ...Installation Reference Manual September 2019 00809 0100 5901 116 Reference Manual ...
Page 204: ...Configuration Reference Manual September 2019 00809 0100 5901 204 Reference Manual ...
Page 208: ...Operation Reference Manual September 2019 00809 0100 5901 208 Reference Manual ...
Page 290: ...Product Certifications Reference Manual September 2019 00809 0100 5901 290 Reference Manual ...