Backing up multiple device configurations using
TankMaster
™
Using Rosemount TankMaster WinSetup to save configuration for multiple devices:
Procedure
1. Start the Rosemount TankMaster WinSetup program.
2. In the
WinSetup
workspace window select the
Devices
folder.
3. Click the right mouse button and choose the
Save Database of All to Files
option.
This option is also available from the
Service/Devices
menu.
4. Select a device from the
Available Devices
pane and press the
Add
button in order to
move it to the
Selected Devices
pane. Repeat for all devices you wish to include.
5. Choose the
Holding Registers
and
Predefined Registers
options (the User-Defined
option should only be used for advanced service).
6. Click the
Browse
button, select a folder and type a name for the backup file.
7. Click the
Start
button to save the database backup.
The backup file can be viewed as a text file in any word processing program:
Reference Manual
Service and troubleshooting
00809-0100-5901
September 2019
Rosemount 5900C Radar Level Gauge
213
Summary of Contents for Rosemount 5900C
Page 1: ...Reference Manual 00809 0100 5901 Rev DA September 2019 Rosemount 5900C Radar Level Gauge ...
Page 6: ...Contents Reference Manual September 2019 00809 0100 5901 vi Reference Manual ...
Page 12: ...Introduction Reference Manual September 2019 00809 0100 5901 12 Reference Manual ...
Page 116: ...Installation Reference Manual September 2019 00809 0100 5901 116 Reference Manual ...
Page 204: ...Configuration Reference Manual September 2019 00809 0100 5901 204 Reference Manual ...
Page 208: ...Operation Reference Manual September 2019 00809 0100 5901 208 Reference Manual ...
Page 290: ...Product Certifications Reference Manual September 2019 00809 0100 5901 290 Reference Manual ...