To open the
Empty Tank Handling
window:
Procedure
1. In the
TankMaster WinSetup
workspace, click the right mouse button on the icon
that represents the desired Rosemount 5900C Radar Level Gauge.
2. From the popup menu choose the
Properties
option.
The
RLG Properties
window appears.
3. In the
RLG Properties
window, select the
Advanced Configuration
tab.
4. Click the
Empty Tank Handling
button:
Figure 4-5: The WinSetup Empty Tank Handling Window
Level alarm is not set when tank is empty
In case the product surface echo is lost in the Empty Tank Detection Area close to the tank
bottom, the device will enter Empty Tank state and an Invalid Level Alarm is triggered
(appears in the
Diagnostics
window).
Enable this check box if you don’t want this alarm to be triggered when the gauge enters
empty tank state.
Activate Extra Echo function
The Extra Echo Detection function is used for tanks with a dome or conical bottom shape
provided the tank bottom does not produce a strong echo when the tank is empty. This
function results in more robust measurements near the tank bottom.
For tanks with a conical bottom, an echo may appear beneath the actual tank bottom
when the tank is empty. If the device is not able to detect the tank bottom this function
can be used to ensure that the device stays in empty tank state as long as this extra echo is
present.
Reference Manual
Configuration
00809-0100-5901
September 2019
Rosemount 5900C Radar Level Gauge
129
Summary of Contents for Rosemount 5900C
Page 1: ...Reference Manual 00809 0100 5901 Rev DA September 2019 Rosemount 5900C Radar Level Gauge ...
Page 6: ...Contents Reference Manual September 2019 00809 0100 5901 vi Reference Manual ...
Page 12: ...Introduction Reference Manual September 2019 00809 0100 5901 12 Reference Manual ...
Page 116: ...Installation Reference Manual September 2019 00809 0100 5901 116 Reference Manual ...
Page 204: ...Configuration Reference Manual September 2019 00809 0100 5901 204 Reference Manual ...
Page 208: ...Operation Reference Manual September 2019 00809 0100 5901 208 Reference Manual ...
Page 290: ...Product Certifications Reference Manual September 2019 00809 0100 5901 290 Reference Manual ...