Calibrate
Prerequisites
Ensure that there is no liquid above the calibration ring
at the end of the still-pipe when
calibrating the gauge. Then the calibration ring is the only object that will be detected by
the gauge. The product level presented by the Rosemount 5900C will be equal to the
position of the calibration ring measured from the Zero Level near the bottom of the tank.
Procedure
1. Check the distance measured by the Rosemount 5900C from the Gauge Reference
to the calibration ring.
This is referred to as the Ullage
value defined by: Ullage = R - L, where
•
R
is the tank height measured from the Tank Reference Point to the Zero Level.
For LPG tanks the calibration ring is used as Zero Level and the Tank Reference
Point is equal to the Gauge Reference Point.
•
L
is the product level measured from the Zero Level.
If the Ullage value is not equal to the actual distance between the
Gauge Reference
Point
and the calibration ring, you will need to adjust the
Calibration Distance
parameter.
2. Right-click the device icon and select the
Properties
→
Geometry
tab.
(10)
See
.
(11)
See
Reference Manual
Configuration
00809-0100-5901
September 2019
Rosemount 5900C Radar Level Gauge
143
Summary of Contents for Rosemount 5900C
Page 1: ...Reference Manual 00809 0100 5901 Rev DA September 2019 Rosemount 5900C Radar Level Gauge ...
Page 6: ...Contents Reference Manual September 2019 00809 0100 5901 vi Reference Manual ...
Page 12: ...Introduction Reference Manual September 2019 00809 0100 5901 12 Reference Manual ...
Page 116: ...Installation Reference Manual September 2019 00809 0100 5901 116 Reference Manual ...
Page 204: ...Configuration Reference Manual September 2019 00809 0100 5901 204 Reference Manual ...
Page 208: ...Operation Reference Manual September 2019 00809 0100 5901 208 Reference Manual ...
Page 290: ...Product Certifications Reference Manual September 2019 00809 0100 5901 290 Reference Manual ...