16. Verify pin position:
a) Open the
Verify Pins
tab.
b) Click the
Pin Verification
button to start the verification process.
c) Compare
Measured Position
with
Nominal Position
(actual position of the
Verification Pin in the Still-pipe).
d) In case the Measured Position deviates from the Nominal Position, note the
Measured Position and return to the
Pin Setup
tab.
e) Enter the Measured position into the
Nominal Position
field and click the
Send
button.
f) Repeat
. until the message
Successful Verification
appears, indicating that Nominal Position corresponds to the Measured
Position.
F
OUNDATION
Fieldbus parameters:
TRANSDUCER 1500>LPG_VER_PIN1_
TRANSDUCER 1500>LPG_PIN1_CONFIGURATION
Reference Manual
Configuration
00809-0100-5901
September 2019
Rosemount 5900C Radar Level Gauge
201