6.2.5
Upgrading the device firmware using TankMaster
™
Rosemount TankMaster WinSetup includes the option to upgrade the Rosemount 5900C
and other devices in a Rosemount Tank Gauging system with new firmware.
Prerequisites
Note
The Rosemount 5900C must not be in SIL Safety mode when it is reprogrammed. Ensure
that necessary safety precautions are considered.
Procedure
1. Ensure that the Rosemount 5900C communicates with TankMaster without any
interruptions or disturbances.
2. In the
Rosemount TankMaster WinSetup
workspace window (Logical View), open
the
Devices
folder and select the device to be upgraded (or select the
Devices
folder to allow multiple devices programming).
3. Right-click and select the
Program
option (
Program All
option for multiple devices
programming). The device will automatically appear in the
Program These Devices
pane.
4. In case the
Devices
folder in the WinSetup workspace was selected for multiple
programming, choose the desired device to be programmed from the
Available
Devices
pane and click the
Move
button.
Service and troubleshooting
Reference Manual
September 2019
00809-0100-5901
216
Reference Manual
Summary of Contents for Rosemount 5900C
Page 1: ...Reference Manual 00809 0100 5901 Rev DA September 2019 Rosemount 5900C Radar Level Gauge ...
Page 6: ...Contents Reference Manual September 2019 00809 0100 5901 vi Reference Manual ...
Page 12: ...Introduction Reference Manual September 2019 00809 0100 5901 12 Reference Manual ...
Page 116: ...Installation Reference Manual September 2019 00809 0100 5901 116 Reference Manual ...
Page 204: ...Configuration Reference Manual September 2019 00809 0100 5901 204 Reference Manual ...
Page 208: ...Operation Reference Manual September 2019 00809 0100 5901 208 Reference Manual ...
Page 290: ...Product Certifications Reference Manual September 2019 00809 0100 5901 290 Reference Manual ...