
4
|
Components
Service manual
244
EBLA04~08E + EDLA04~08E
Daikin Altherma 3 M
ESIE22-09 – 2022.05
2
Route the wiring harness towards the water pressure sensor and connect the
wiring harness to the water pressure sensor.
3
Fix the wiring harness using new tie straps.
Is the problem solved?
Action
Yes
No further actions required.
No
Return to the troubleshooting of the
specific error and continue with the
next procedure.
To remove the water pressure sensor
Prerequisite:
Stop the unit operation via the user interface.
Prerequisite:
Turn OFF the respective circuit breaker.
Prerequisite:
Remove the required plate work, see
202].
1
Drain water from the water circuit, see
269].
2
Disconnect the water pressure sensor connector from the water pressure
sensor.
a
b
a
Water pressure sensor harness
b
Water pressure sensor
3
Screw the water pressure sensor out of the coupling piece.
4
To install the water pressure sensor, see
To install the water pressure sensor
1
Check that a new O-ring is installed on the water pressure sensor.
2
Screw the water pressure sensor in the coupling piece.
draft – 31/05/2022 15:53
Summary of Contents for EBLA04E V Series
Page 322: ...14 14...
Page 323: ......