22
23
ENGLISH
11
Electro-magnetic compatibility
Recommended separation between portable RF communication equipment mobiles, and the
« Xenon light source » reference equipment
The medical device is intended for use in an electromagnetic environment in which interference due
to RF radiation is controlled. Do not to use portable RF communication devices (including devices
such as antenna cables and external antennas) closer than 30 cm (12 inches) from any part of the
S218 (or S230) device; this includes cables specified by the manufacturer. Otherwise the performance
of these devices could be affected.
Maximum assigned
output
power of the trans-
mitter
W
Separation distance as a function of the transmitter’s frequency
M
150 kHz to 80
MHz
80 MHz to 800
MHz
800 MHz to 2.7 GHz
d = 1,16 √
P
d = 1,16 √
P
d = 2,33 √
P
0.01
0.1
1
10
100
0.116
0.366
1.16
3.66
11.6
0.116
0.366
1.16
3.66
11.6
0.233
0.736
2.33
7.36
23.3
For transmitters of which the maximum output power is not listed above, the recommended
separation distance d in meters (m) can be established by using the equation applicable for the
transmitter frequency, in which
P is the transmitter’s maximum output power in Watts (W) assigned
by the transmitter’s manufacturer.
NOTE 1 :
These recommendations may not apply in all situations. The propagation of electromagnetic
waves is modified by absorption and reflection due to structures, objects and people.
Summary of Contents for SOPRO 218
Page 2: ...2...
Page 4: ...4...
Page 5: ...INTUITIVE CONNECTION English 5 ENGLISH...
Page 6: ...6 ENGLISH...
Page 16: ...16 ENGLISH 8 Lamp change Correspondence 1 3 5 2 4 6...
Page 24: ...24 ENGLISH 12 Symbols L L C2 S1 S1 C1 C1 S2 S2 S3 S3 S4 S4...
Page 25: ...25 ENGLISH...
Page 26: ...26...
Page 27: ...INTUITIVE CONNECTION Fran ais 27 FRAN AIS...
Page 28: ...28 FRAN AIS...
Page 38: ...38 FRAN AIS 8 Changement de la lampe Correspondance 1 3 5 2 4 6...
Page 46: ...46 FRAN AIS 12 Symboles L L C2 S1 S1 C1 C1 S2 S2 S3 S3 S4 S4...
Page 47: ...47 FRAN AIS...
Page 48: ......
Page 49: ...INTUITIVE CONNECTION Espa ol 49 ESPA OL...
Page 50: ...50 ESPA OL...
Page 60: ...60 ESPA OL 8 Cambio de l mpara Correspondencia 1 3 5 2 4 6...
Page 68: ...68 ESPA OL 12 S mbolos L L C2 S1 S1 C1 C1 S2 S2 S3 S3 S4 S4...
Page 69: ...69 ESPA OL...
Page 70: ......
Page 71: ...INTUITIVE CONNECTION Deutsch 71 DEUTSCH...
Page 72: ...72 DEUTSCH...
Page 82: ...82 DEUTSCH 8 Austausch der lampe Korrespondenz 1 3 5 2 4 6...
Page 90: ...90 DEUTSCH 13 Symbole L L C2 S1 S1 C1 C1 S2 S2 S3 S3 S4 S4...
Page 91: ...91 DEUTSCH...
Page 92: ......
Page 93: ...INTUITIVE CONNECTION Italiano 93 ITALIANO...
Page 94: ...94 ITALIANO...
Page 104: ...104 ITALIANO 8 Sostituzione della lampadina corrispondenza 1 3 5 2 4 6...
Page 112: ...112 ITALIANO 12 Simboli L L C2 S1 S1 C1 C1 S2 S2 S3 S3 S4 S4...
Page 113: ...113 ITALIANO...