Müse Parameter Descriptions
286
Respiratory Gain Factor
The
Respiratory Gain Factor
determines how strong an influence arterial CO
2
levels have on the
simulated patient’s tidal volume and respiratory rate. Under default conditions (value = 1), when
arterial CO
2
levels rise, the patient’s respiratory rate and
tidal volume show a transitory increase in an
attempt to return the patient to the physiological control CO
2
set-point. If the
Respiratory Gain Factor
is increased to more than 1, the patient has a more pronounced response, while values less than 1
correspond to a blunted response.
Default
: 1
Range
: 0.00 - 10.00
Respiratory Quotient
Respiratory Quotient
is the rate of carbon dioxide production divided by the rate of oxygen
consumption. Changes to the
Respiratory Quotient
parameter alter the rate of carbon dioxide
production relative to the rate of oxygen consumption.
Default
: 0.8
Range
: 0.70 - 1.10
Volume/Rate Control Factor
Ventilatory responses to increased arterial carbon dioxide or decreased arterial oxygen may take the
form of increased tidal volume, increased respiratory rate, or both. The
Volume/Rate Control Factor
determines these relative changes. At a value of 1, increased and decreased ventilatory drive affect tidal
volume and respiratory rate equally. When the
Volume/Rate Control Factor
is greater than 1,
increased or decreased minute ventilation is predominantly achieved by changes in tidal volume.
When the
Volume/Rate Control Factor
is less than 1, ventilatory changes are affected primarily by
changes in
respiratory rate.
For example, set the
Volume/Rate Control Factor
to 0.1 and increase the shunt fraction to 0.4 to
decrease the arterial O
2
. The patient responds to falling arterial oxygen levels with increased minute
ventilation. Increasing the
respiratory rate with minimal increase in tidal volume produces this.
Default
: 1
Range
: 0.1 - 10.0
Chest Wall Capacity
The
Chest Wall Capacity
parameter sets the total (combined) intrapleural and lung volumes at which
the chest wall is considered distended. Also, see
Chest Wall Compliance Factor
and
Distended Chest
Wall Compliance Factor
.
Default
: 3900
Range
: 1500 - 3900
Summary of Contents for Fidelis Lucina
Page 1: ...User Guide Maternal Fetal Simulator...
Page 2: ...2015 CAE Healthcare 905K470052 v2 1...
Page 28: ...Table Of Contents xxvi This page intentionally blank...
Page 36: ...Equipment Overview 8 This page intentionally blank...
Page 190: ...Using M se 162 This page intentionally blank...
Page 226: ...Using the TouchPro CTG Monitor 198 This page intentionally blank...
Page 266: ...Using Lucina 238 This page intentionally blank...
Page 298: ...Care and Maintenance 270 This page intentionally blank...
Page 389: ......