Müse Parameter Descriptions
297
Pacing Current
The
Pacing Current
parameter is used to simulate a specified amount of current discharged via an
external cardiac pacer. Setting this parameter results in the characteristic pacing signal on the ECG
waveform when the pacing current is at or above the capture threshold. Also, see
Pacing Capture
Threshold
.
Default
: 0 mA
Range
: 0 mA - 200 mA
Pacing Rate
The
Pacing Rate
parameter determines the cardiac rate (in beats/minute) when the pacing current is at
or above the pacing capture threshold. Also, see
Pacing Current
and
Pacing Capture Threshold
.
Default
: 80 beats per minute
Range
: 0 beats per minute - 119 beats per minute
Pacing Capture Threshold
The
Pacing Capture Threshold
parameter determines the minimum pacing current necessary to pace
the heart via an external cardiac pacer. Also see
Pacing Current
. Pacing current values below the
pacing capture threshold have no effect on the patient’s heart rate.
Default
: 50 mA
Range
: 0 mA - 119 mA
Cold Fluid Inject
The Cold Fluid Inject parameter is used to simulate the injection of 10 mL saline into the pulmonary
artery (PA) catheter. The appropriate Thermodilution waveform and cardiac output measurement are
then displayed on the Patient Status Display or TouchPro software.
Summary of Contents for Fidelis Lucina
Page 1: ...User Guide Maternal Fetal Simulator...
Page 2: ...2015 CAE Healthcare 905K470052 v2 1...
Page 28: ...Table Of Contents xxvi This page intentionally blank...
Page 36: ...Equipment Overview 8 This page intentionally blank...
Page 190: ...Using M se 162 This page intentionally blank...
Page 226: ...Using the TouchPro CTG Monitor 198 This page intentionally blank...
Page 266: ...Using Lucina 238 This page intentionally blank...
Page 298: ...Care and Maintenance 270 This page intentionally blank...
Page 389: ......