Müse Parameter Descriptions
302
Venous Return Resistance Factor
The
Venous Return Resistance Factor
parameter adjusts the resistance between the extrathoracic
and intrathoracic venous compartments. Raising the value increases the resistance, while lowering the
value decreases the resistance.
With less blood returning to the heart, there is a reduced volume entering the ventricles prior to
ventricular contraction. This results in a drop in the cardiac output and decrease in arterial blood
pressures. The heart rate increases due to feedback from the physiological control mechanisms in an
attempt to maintain adequate blood pressures.
Default
: 1
Range
: 0.10 - 100.00
Baroreceptor Gain (Overall) Factor
The
Baroreceptor Gain (Overall) Factor
parameter adjusts the influence of mean arterial pressure
(MAP) on heart rate, contractility, systemic vascular resistance and venous capacity. Use this parameter
to adjust how vigorously the heart and vasculature respond to blood pressure changes. The degree of
increase in heart rate or vascular response is influenced by the baroreceptor gain (overall) factor.
For example, when blood pressure falls, the heart rate increases, the arteries increase their vascular
tone (resistance) and there is less pooling of the blood in the venous system, all in an attempt to
maintain adequate blood pressure. A baroreceptor gain (overall) factor value of less than 1 corresponds
to baroreceptor depression. A baroreceptor gain (overall) factor value greater than 1 leads to a stronger
response to MAP changes.
Default
: 1
Range
: 0.00 - 100.00
Baroreceptor Gain (Cardiac) Factor
The
Baroreceptor Gain (Cardiac) Factor
parameter selectively adjusts the influence of mean arterial
pressure (MAP) on the heart rate and contractility, influencing how much the heart rate increases or
decreases with changes in blood pressure. Use this parameter to adjust how vigorously the heart
responds to blood pressure changes.
A baroreceptor gain (cardiac) factor of less than 1 corresponds to baroreflex depression (e.g., less heart
rate response to MAP changes). A value greater than 1 leads to a stronger response to MAP changes.
Default
: 1
Range
: 0.00 - 10.00
Summary of Contents for Fidelis Lucina
Page 1: ...User Guide Maternal Fetal Simulator...
Page 2: ...2015 CAE Healthcare 905K470052 v2 1...
Page 28: ...Table Of Contents xxvi This page intentionally blank...
Page 36: ...Equipment Overview 8 This page intentionally blank...
Page 190: ...Using M se 162 This page intentionally blank...
Page 226: ...Using the TouchPro CTG Monitor 198 This page intentionally blank...
Page 266: ...Using Lucina 238 This page intentionally blank...
Page 298: ...Care and Maintenance 270 This page intentionally blank...
Page 389: ......