Müse Parameter Descriptions
288
Venous CO
2
Shift
The
Venous CO
2
Shift
parameter affects the partial pressure of CO
2
in the venous blood. Changing this
parameter allows large and rapid shifts in total body CO
2
concentration. Increases in alveolar and
arterial CO
2
follow rapidly in a physiologically correct magnitude and time course.
This parameter is useful for giving a “bolus” of CO
2
to the venous system. The alveolar and arterial CO
2
levels rise rapidly in response to the added carbon dioxide but soon return to “pre-bolus” levels as
increased ventilation efforts work to eliminate the added CO
2
. Therefore, the rise in CO
2
levels is only
transitory. This parameter can be used to simulate external CO
2
administration such as that used during
laparoscopy.
Default
: 0 mmHg
Range
: 0.0 mmHg - 60.0 mmHg
Bronchial Resistance Factor (Left and Right)
The
Bronchial Resistance Factor
parameter can be used to set the rate of left and right bronchial
resistance individually. The rate of resistance can also be set to occur over time.
Default:
1
Range:
0.00 - 250,000.00
Alveolar Enflurane
The
Alveolar Enflurane
parameter is used to simulate the presence of enflurane in the alveolar space
without using real anesthetic vapors. The enflurane percentage is input to the drug models to achieve
the expected pharmacodynamic effects (e.g., respiratory depression).
Using this parameter bypasses pharmacokinetics, real and instructor-specified inspired fractions,
venous content, lung perfusion and ventilation. This parameter can be used to focus on
cardiorespiratory effects (pharmacodynamics).
NOTE:
Müse does not verify a 100 % sum of all fractions, because this would require all fractions to be
set.
Default
: Modeled
Range
: 0.00% - 5.00%
Summary of Contents for Fidelis Lucina
Page 1: ...User Guide Maternal Fetal Simulator...
Page 2: ...2015 CAE Healthcare 905K470052 v2 1...
Page 28: ...Table Of Contents xxvi This page intentionally blank...
Page 36: ...Equipment Overview 8 This page intentionally blank...
Page 190: ...Using M se 162 This page intentionally blank...
Page 226: ...Using the TouchPro CTG Monitor 198 This page intentionally blank...
Page 266: ...Using Lucina 238 This page intentionally blank...
Page 298: ...Care and Maintenance 270 This page intentionally blank...
Page 389: ......