Care and Maintenance
251
2. Rinse thoroughly the trauma fill tank of any simulated blood.
3. Fill with distilled water or the 50/50 mixture of distilled water and white
vinegar.
4. Connect the trauma fill tank to the mannequin. The blue tube connects to
the
BLOOD FILL
port and yellow tube connects to the
BLOOD VENT
port.
5. Pressurize the trauma fill tank using the hand pump. Wait until there is
reflow in the reclaim bottle.
6. Place absorbent pads under the buttocks and the vaginal opening.
7. From the Instructor Workstation, click on the
System Administration
icon
in the upper-right corner of the Müse home screen.
The System Administration Button
The System Administration screen appears.
8. Click on the
Maintenance
tab.
The Maintenance Panel
The Maintenance panel appears.
9. Click the
Flush System
button.
The Flush in progress window appears.
10. Run until the fluids run clear from the simulator and click the
Finished
button.
11. Disconnect the trauma fill tank from the mannequin and depressurize.
The
Maintenance
panel
The
Flush
System
button
The
Maintenance
tab
Summary of Contents for Fidelis Lucina
Page 1: ...User Guide Maternal Fetal Simulator...
Page 2: ...2015 CAE Healthcare 905K470052 v2 1...
Page 28: ...Table Of Contents xxvi This page intentionally blank...
Page 36: ...Equipment Overview 8 This page intentionally blank...
Page 190: ...Using M se 162 This page intentionally blank...
Page 226: ...Using the TouchPro CTG Monitor 198 This page intentionally blank...
Page 266: ...Using Lucina 238 This page intentionally blank...
Page 298: ...Care and Maintenance 270 This page intentionally blank...
Page 389: ......